The HiTemp140-FP is a high temperature data logger built for use in harsh environments. The body of the HiTemp140-FP can withstand temperatures ranging from -40 °C to +140 °C (-40 °F to +284 °F), but can extend to higher temperatures when used with MadgeTech’s thermal barrier options.
The HiTemp140-FP is compatible with the latest MadgeTech 4 Software making starting, stopping, and downloading simple and easy. Graphical, tabular, and summary data is provided for analysis and can be viewed as customizable engineering units.
Memory | 32,356 readings |
Reading Rate | 1 reading every 0.25 seconds up to 1 reading every 24 hours |
Baud Rate | 125,000 |
Battery Life | 2 years typical (1 minute reading rate) Fast reading intervals and extreme temperatures will shorten the battery life substantially User Replaceable Battery |
Material | Body: 316 Stainless Steel, Radel Probe: PFA Insulated Cable with Stainless Steel Tip |
Dimensions | Body: 2.95 in x 0.97 in x 0.97 in (75 mm x 24.6 mm x 24.6 mm) Probes: HiTemp140-FPST-6: 6 in x 0.13 in (152.5 mm x 3.2 mm) HiTemp140-FPST-12: 12 in x 0.13 in (304.8 mm x 3.2 mm) HiTemp140-FPST-36: 36 in x 0.13 in (914 mm x 3.2 mm) HiTemp140-FPST-72: 72 in x 0.13 in (1829 mm x 3.2 mm) |
Key Ring End Cap | Key ring replaces standard bottom Thru-hole: 0.150″ |
IP Rating | IP68 |
Submersible | Yes |
Operating Environment | -40 °C to +140 °C (-40 °F to +284 °F) Body of Logger Cannot Exceed 140 °C 0 %RH to 100 %RH, 0 PSIA to 100 PSIA With Thermal Shield: -200 °C to +250 °C (-328 °F to +482 °F) |
Required Interface Package | IFC400 or IFC406 |
Approvals | CE |
Weight | 0.1875 LBS |
Temperature Sensor | Flexible RTD Probe |
Temperature Resolution | 0.01 °C (0.018 °F) |
Temperature Calibrated Accuracy | ±0.1 °C (0.18 °F) |
Probe Temperature Range | -60 °C to +260 °C (-76 °F to +500 °F) |
Image | Probe Type | Material | Operating Environment | Probe Measurement Range |
---|---|---|---|---|
Rigid | Metal | Temperature: -40 °C to +140 °C (-40 °F to +284 °F) Humidity: 0 %RH to 100 %RH Pressure: 0.002 PSIA to 100 PSIA |
Temperature: -200 °C to +260 °C (-328 °F to +500 °F) Probe only. Body must not exceed Operating Environment |
|
Rigid Transitional Diameter (TD) | Metal | Temperature: -40 °C to +140 °C (-40 °F to +284 °F) Humidity: 0 %RH to 100 %RH Pressure: 0.002 PSIA to 100 PSIA |
Temperature: -200 °C to +260 °C (-328 °F to +500 °F) Probe only. Body must not exceed Operating Environment |
|
Flexible (FP) | Bendable PTFE | Temperature: -40 °C to +140 °C (-40 °F to +284 °F) Humidity: 0 %RH to 100 %RH Pressure: 0.002 PSIA to 60 PSIA |
Temperature: -60 °C to +260 °C (-76 °F to 500 °F) Probe only. Body must not exceed Operating Environment |
|
PT | Bendable Metal | Temperature: -40 °C to +140 °C (-40 °F to +284 °F) Humidity: 0 %RH to 100 %RH Pressure: 0.002 PSIA to 100 PSIA |
Temperature: -200 °C to +350 °C (-328 °F to +662 °F) Probe only. Body must not exceed Operating Environment |
|
Fast Response (FR) | Metal | Temperature: -40 °C to +140 °C (-40 °F to +284 °F) Humidity: 0 %RH to 100 %RH Pressure: 0.002 PSIA to 100 PSIA |
Temperature: -200 °C to +260 °C (-328 °F to +500 °F) Probe only. Body must not exceed Operating Environment |