The HiTemp140 is a submersible, high temperature data logger built for use in harsh environments. This device is extremely versatile and can be tailored to the application with various probe lengths and styles available.
The HiTemp140 can withstand temperatures ranging from -40 °C to +140 °C (-40 °F to +284 °F), but is capable of measuring extended temperatures up to 260 °C (500 °F) (probe dependent). With a storage capacity of up to 32,256 time and date stamped readings, the HiTemp140 can be programmed for immediate start or delayed start up to 18 months in advance.
The HiTemp140 is compatible with the latest MadgeTech 4 Software making starting, stopping, and downloading simple and easy. Graphical, tabular, and summary data is provided for analysis and can be viewed as customizable engineering units.
Body of data logger cannot exceed 140 °C (284 °F), but can extend to higher temperatures when used with MadgeTech’s thermal barrier options.
HiTemp140-FR fast response probe is very thin to increase the response time, hence fragile, should be used carefully to avoid physical damages, which is not covered under manufacturer warranty.
HiTemp140-PT bendable metal probe can be bent to a 0.25 inch bend radius. The probe should not be bent within 1 inch of either weld joint, which will cause physical damages to the weld joints, which is not covered under manufacturer warranty. Where possible, the HiTemp140-FPST flexible probe will be a better alternative.
Memory | 32,256 readings |
Reading Rate | 1 reading every second up to 1 reading every 24 hours |
Baud Rate | 125,000 |
Battery Life | 2 years typical (1 minute reading rate) Fast reading intervals and extreme temperatures will shorten the battery life substantially User Replaceable Battery |
Material | 316 Stainless Steel, Radel |
HiTemp 140 Rigid and Rigid Transitional Diameter Dimensions | HiTemp140-1 Body: 1.55 in x 0.97 in dia. (40 mm x 25 mm dia.) Probe: 1 in x 0.125 in dia. (0.188 in transitional dia.) 25 mm x 3.2 mm dia. (4.8 mm transitional dia.) Body (for 2 in, 5.25 in, 7 in;): 1.75 in x 0.97 in dia. (45 mm x 25 mm dia.)HiTemp140-2 : 2.0 in x 0.188 in dia. (51 mm x 4.8 mm)HiTemp140-2-TD: 2.0 in x 0.125 in dia. (0.188 in transitional dia.) (51 mm x 3.2 mm dia. (4.8 mm transitional dia.)HiTemp140-5.25: 5.25 in x 0.188 in dia. (134 mm x 4.8 mm)HiTemp140-5.25-TD: 5.25 in x 0.125 in dia. (0.188 in transitional dia.) (134 mm x 3.2 mm dia. (4.8 mm transitional dia.)HiTemp140-7: 7.0 in x 0.188 in dia. (178 mm x 4.8 mm dia.) |
HiTemp140 Fast Response Probe Dimensions | Body: 1.89 in x 0.97 in x 0.97 in (48.1 mm x 24.6 mm x 24.6 mm) Probe: 1.75 in x 0.0625 in dia. (0.188 in transitional dia.) (44.5 mm x 1.59 mm (4.78 mm Transitional dia.)) |
Key Ring End Cap | Key ring replaces standard bottom Thru-hole: 0.150″ |
IP Rating | IP68 |
Submersible | Yes |
Operating Environment | -40 °C to +140 °C (-40 °F to +284 °F); 0 %RH to 100 %RH; 0.002 PSIA to 100 PSIA (Body of Logger Cannot Exceed 140 °C) |
Required Interface Package | IFC400 |
Approvals | CE |
Temperature Sensor | 100 Ω Platinum RTD |
Temperature Resolution | 0.01 °C (0.018 °F) |
Temperature Calibrated Accuracy | ±0.1 °C/±0.18 °F (20 °C to +140 °C/68 °F to +284 °F) ±0.3 °C/±0.54 °F (-20 °C to +19.99 °C/-4 °F to +67.98 °F) ±0.4 °C/±0.72 ºF (-40 °C to -20.01 °C/-40 °F to +-4.02 °F) |
Probe Temperature Range | -200 °C to +260 °C (-328 °F to +500 °F) (Body of Logger Cannot Exceed 140 °C) |
Image | Probe Type | Material | Operating Environment | Probe Measurement Range |
---|---|---|---|---|
Rigid | Metal | Temperature: -40 °C to +140 °C (-40 °F to +284 °F) Humidity: 0 %RH to 100 %RH Pressure: 0.002 PSIA to 100 PSIA |
Temperature: -200 °C to +260 °C (-328 °F to +500 °F) Probe only. Body must not exceed Operating Environment |
|
Rigid Transitional Diameter (TD) | Metal | Temperature: -40 °C to +140 °C (-40 °F to +284 °F) Humidity: 0 %RH to 100 %RH Pressure: 0.002 PSIA to 100 PSIA |
Temperature: -200 °C to +260 °C (-328 °F to +500 °F) Probe only. Body must not exceed Operating Environment |
|
Flexible (FP) | Bendable PTFE | Temperature: -40 °C to +140 °C (-40 °F to +284 °F) Humidity: 0 %RH to 100 %RH Pressure: 0.002 PSIA to 60 PSIA |
Temperature: -60 °C to +260 °C (-76 °F to 500 °F) Probe only. Body must not exceed Operating Environment |
|
PT | Bendable Metal | Temperature: -40 °C to +140 °C (-40 °F to +284 °F) Humidity: 0 %RH to 100 %RH Pressure: 0.002 PSIA to 100 PSIA |
Temperature: -200 °C to +350 °C (-328 °F to +662 °F) Probe only. Body must not exceed Operating Environment |
|
Fast Response (FR) | Metal | Temperature: -40 °C to +140 °C (-40 °F to +284 °F) Humidity: 0 %RH to 100 %RH Pressure: 0.002 PSIA to 100 PSIA |
Temperature: -200 °C to +260 °C (-328 °F to +500 °F) Probe only. Body must not exceed Operating Environment |
Complete system for performing autoclave validati…
Complete system for performing autoclave validati…
High temperature data logger with a flexible RTD …