1300 662 720

Customer Support

1300 662 720

Customer Support

1300 662 720

Customer Support

M1220 | Muilti-logger

Multilogger – thermo hygro meter with 2 MiniDIN and 2 Thermocouple inputs

Available on back-order

$ 2,125.00 Ex. GST

Share

Share on facebook
Share on twitter
Share on whatsapp
Share on email

Universal datalogger with 4 inputs. Two inputs for temperature-humidity probes and two inputs for thermocouples K, J, S, B, T, N.
Multilogger can be installed permanently to Ethernet network or work as portable device.
Included is traceable calibration certificate in accordance with EN ISO/IEC17025.

The Multilogger is a battery operated datalogger capable of measuring and recording sensor and electrical values through its Inputs.
Data logger can be connected to Ethernet. Then measured values can be viewed on a web browser, as well as stored and downloaded for later analysis.

Warranty: 3 years

INCLUDED IN DELIVERY

    • M1220
    • Instruction Manual
    • Traceable Calibration Certificate
    • 3x Rechargeable Battery AA
    • Power Supply
    • USB Cable
    • Wall Holder
    • FREE Analytical Software COMET Vision

EXTERNAL PROBES ARE NOT INCLUDED IN DELIVERY.

MEASURED VALUE

Temperature
Relative humidity
Counted values (dew point…)

TEMPERATURE SENSOR – EXTERNAL PROBE PT1000

Measuring range -200 to +600 °C
Accuracy ±0.2 °C (-200 to +100 °C);
±0.2 % from reading from +100 to +600 °C
Resolution 0.1 °C

TEMPERATURE+HUMIDITY SENSOR – EXTERNAL PROBE T+RH DIGI

Measuring range according to the connected probe Digi/M
Accuracy according to the connected probe Digi/M
Resolution 0.1 °C

DEW POINT

Measuring range according to the connected probe Digi/M
Accuracy according to the connected probe Digi/M
Resolution 0.1 °C

TEMPERATURE SENSOR – THERMOCOUPLE K, N

Measuring range -200 to +1300 °C
Accuracy ± 0,3 % from the measured value + 1,5 °C
Resolution 0.1 °C

TEMPERATURE SENSOR – THERMOCOUPLE J

Measuring range -200 to +750 °C
Accuracy ± 0,3 % from the measured value + 1,5 °C
Resolution 0.1 °C

TEMPERATURE SENSOR – THERMOCOUPLE S

Measuring range 0 to +1700 °C
Accuracy ± 0,3 % from the measured value + 1,5 °C
Resolution 0.1 °C

TEMPERATURE SENSOR – THERMOCOUPLE B

Measuring range 100 to +1800 °C
Accuracy ± 0,3 % from the measured value + 1 °C in the range +300 to +1800 °C
Resolution 0.1 °C

TEMPERATURE SENSOR – THERMOCOUPLE T

Measuring range -200 to +400 °C
Accuracy ± 0,3 % from the measured value + 1,5 °C
Resolution 0.1 °C
DC VOLTAGE
Measuring range -60 to +140 mV; -18 to +18 mV
Accuracy ±100 µV; ±20 µV
Resolution 0,1 µV

GENERAL TECHNICAL DATA

Operating temperature -10 to +60 °C
Channels 4x external temperature probe
Counted values dew point, absolute humidity, specific humidity, mixing ratio, specific enthalpy
Memory 1 000 000 values (noncyclic record), 600 000 values (cyclic record)
Recording interval 1s to 24 hours
Display and alarm refresh each 10 s
Recording mode noncyclic – data logging stops after filling the memory
cyclic – after filling memory oldest data is overwritten by new
Available temperature units degrees Celsius, degrees Fahrenheit
Communication interface Ethernet, USB, RS232
Ethernet communication protocols WWW, SNMPv1, XML, DATALINK
Alarm protocols E-Mail
Power AC adapter 230Vac/5Vdc (neccessary for Ethernet interface)
Battery life 3 – 4 weeks
Protection class IP20
Dimensions (without connectors) 178 x 95 x 37mm, without attached cables
Weight (including batteries) approx. 380 g
Warranty 3 years

Up to 4 physical Inputs – 2x MiniDin + 2x Thermocouple

Multilogger M1220 inputs
  • Temperature probes Pt1000 with built-in configuration memory (serie xxx/M)
  • Temperature/relative humidity probes with digital output – DigiS/M, DigiL/M, DIGIH/M
  • The probes are interchangeable without calibration to a specific device and regardless of the length of the cable. The length of the cable can be 1, 2, 5, 10, 15 meters.
  • Themocouple probe K, J, S, B, T, N
  • Direct input of bipolar voltages with range up to -60 to +140 mV
  • Sensors with woltage output (range -60 to +140mV; accuracy ±100 µV or -18 to +18mV; accuracy ±20 µV)

Evaluation of up to 16 variables

Multilogger - LCD detail

Up to 16 variables may be calculated from the four connected probes / sensors. This is the sum of the measured and calculated values. The calculated values can be a further expression of moisture – dew point temperature, absolute humidity, specific humidity, mixing ratio, specific enthalpy. It can be also the result of inter-channel conversions, e.g. the difference of two connected temperature probes.

The measured values are stored in the internal non-volatile memory. Memory capacity is up to 1 million values. Logging mode can be adjusted as non-cyclic, when logging stops after filling the memory, or cyclic, when after fulfilling the memory oldest stored values are overwritten by new ones.

Measured values are displayed on the screen with backlight option and the ability to change the size of display digits

Power supply

Multilogger - Power Supply
Multilogger - Power Supply

Multilogger can be powered by AC adapter or by rechargeable batteries with charging capability. Batteries are always a part of the device (you can also use standard alkaline AA batteries). Batteries power the device for several months.

While using the AC adapter, the batteries provide backup operations and provide recording and measurement during power failure. Thanks to the Ethernet output the device informs about the failure via e-mail message.

Alarm limits and alarm indication

Multilogger - Alarm limits

It is possible to set two independent alarm limits for each channel (i.e. measured or calculated value).

Alarm signalization can be acoustic (built-in beeper), optical, sending an e-mail alert or an alarm output. The alarm output is for connecting an external audio or visual alarm.

Communication interfaces

Multilogger is always equipped with USB connector for communication, data analysis and also for charging of the rechargeable batteries.

Multilogger - Alarm limits
  • send an e-mail when an alarm state starts or ends
  • use DATALINK: display current values or download values from memory to COMET Database in your PC. Software COMET Database contains many useful tools for data analysis – graphs, tables, statistics, etc.
  • simple wireless communication with Wi-Fi TP-LINK router
  • third-party applications to read the actual measured values using universal protocols SNMPv1 and XML
  • view current measured values using your web browser

Output ALARM OUT

Multilogger - Alarm limits and alarm indication
  • Output can be used for controling other devices when alarm is launched, such as external buzzer, telephone dialer etc.
  • Type of output – open collector transistor
  • Max. switching current – 100 mA
  • Max. voltage on output – 24 V
  • Auxiliary voltage at terminal – +5 V (only when the mains supply is present)
  • Users can receive alarms if the values go outside the paramters set. Alarms are indicated through an internal buzzer , 3 LEDs or email alerts via Ethernet port.

Instruction Manual

Quick Start Manual

Declaration of Comformity

MIB

 

CATALOG – Multilogger