1300 662 720

Customer Support

1300 662 720

Customer Support

1300 662 720

Customer Support

D3121P | Compressed Air

Data logging  thermometer, hygrometer for compressed air up to 25 bars

Available on back-order

  • Please select the model that you require. Base price shown, price will change based on the selection.

    • 0 $
    • 47.5 $
    • 58.5 $

$ 1,108.33 Ex. GST

Share

Share on facebook
Share on twitter
Share on whatsapp
Share on email

Thermometer, hygrometer is designed for measurement and data logging of temperature, relative humidity for compressed air up to 25 bars with 1 m cable into the internal non-volatile memory.

Free PC program enables to transfer recorded data to the computer for evaluation. LCD back-light.

Data logger is connected to the PC through USB cable only for data transfer from the memory. During connection with the PC no measurement and record is enabled.

Simultaneous reading of temperature and relative humidity, dew point temperature reading selectable, external temperature-humidity probe with 1 meter cable. Cable lengths 2 and 4 meters available optionally. Probe diameter is 18 mm, length 135 mm.

INCLUDED IN DELIVERY:
  • D3121P
  • Instruction Manual
  • Traceable Calibration Certificate
  • Battery 9 V
  • Mini USB Cable
  • Plastic Transport Case
  • Self Adhesive Dual Lock for Easy Installation
  • FREE Analytical Software COMET Vision
Warranty: 3 years

MEASURED VALUE

Temperature
Relative humidity

TEMPERATURE SENSOR

Measuring range -30 to +105 °C
Accuracy ±0.4 °C
Resolution 0.1 °C

HUMIDITY SENSOR

Measuring range 0 to 100 % RH
Accuracy ±2.5 % RH from 5 to 95 % at 23 °C
Resolution 0.1% RH

DEW POINT

Measuring range -50 to +80 °C
Accuracy ±1.5 °C at ambient temperature T <25 °C and RH >30 %
Resolution 0.1 °C

GENERAL TECHNICAL DATA

Operating temperature electronics -10 to +60 °C, external probe -30 to +105 °C
Channels 1x connectable temperature+humidity probe
Acoustic signalization + backlit LCD YES
Memory 16,000 values (non-cyclic record)
Recording interval adjustable from 10 s to 24 h
Measuring interval and display reading refresh 0.7 to 5 s
Recording mode noncyclic – data logging stops after filling the memory
cyclic – after filling memory oldest data is overwritten by new
Real time clock year, leap year, month, day, hour, minute, second
Power battery 9 V or 12 Vss adaptor with NiMH 9 V battery
Battery life approx. 6 months
Protection class IP20 electronics, IP40 sensors
Dimensions (without connectors) 191 x 71 x 27 mm; length/diameter of external probe 135/18 mm
External probe cable length 1 meter
Weight (including batteries) approx. 150 g
Warranty 3 years

D3121P Instruction Manual

Quick Start Manual